Published March 2010 | Version v1
Journal article

Measurement of Cerenkov light in a fiber-optic radiation sensor by using high-energy photon and electron beams

  • 1. Konkuk University, Chungju (Korea, Republic of)
  • 2. Catholic University, Daegu (Korea, Republic of)
  • 3. Soonchunhyang University, Asan (Korea, Republic of)
  • 4. Dongguk University, Gyeongju (Korea, Republic of)
  • 5. Cheju National University, Cheju (Korea, Republic of)

Description

In this study, we used a charge coupled device to measure scintillating and Cerenkov light generated in a scintillating fiber-optic radiation sensor irradiated by high-energy photon and electron beams. The intensities of Cerenkov light are measured and characterized as a function of the incident angles of the high-energy photon and electron beams from a clinical linear accelerator. To minimize or remove Cerenkov light, we investigated a subtraction method using a dummy optical fiber and a wavelength discrimination method using optical filters. Also, the intensities of Cerenkov light induced by high-energy photon and electron beams are compared.

Additional details

Publishing Information

Journal Title
Journal of the Korean Physical Society
Journal Volume
56
Journal Issue
3
Series
19 refs, 5 figs
Journal Page Range
p. 765-769
ISSN