Approximation of collimator scatter factory, Sc, by a saturation model
Creators
- 1. Niigata Univ., School of Health Sciences, Niigata, Niigata (Japan)
- 2. Niigata Univ., Hospital, Niigata, Niigata (Japan)
- 3. Tsuruoka Municipal Shonai Hospital, Tsuruoka, Yamagata (Japan)
Description
To more easily estimate accurate values of collimator scatter factor, Sc, we suggest a two-component saturation model that accounts for scatter from the primary collimator and flattening filter and from the collimator jaws. This model, which assumes an exponential distribution of scatter intensity, was tested by in-air measurements using a mini-phantom for 4 MV and 10 MV X-rays of a Clinac 2100 C/D linear accelerator. The results showed a good fit of this model to our measured data (R2>0.9993). When the measured value was divided into the primary collimator/flattening filter component and the collimator jaw component, as expected, the former component showed a rapid and full saturation curve with increased field size, while the latter showed an almost linearly increasing curve. Therefore, we think that this saturation model is useful for the estimation of Sc and is applicable to monitor unit calculation for an asymmetric field. (author)
Additional details
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi
- Journal Volume
- 62
- Journal Issue
- 12
- Journal Page Range
- p. 1675-1681
- ISSN
- 0369-4305
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 38044293
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- APPROXIMATIONS; BACKSCATTERING; COLLIMATORS; FILTERS; LINEAR ACCELERATORS; MATHEMATICAL MODELS; PHANTOMS; RADIATION DOSES; RADIATION SCATTERING ANALYSIS; RADIOTHERAPY; SATURATION
- Descriptors DEC
- ACCELERATORS; CALCULATION METHODS; CHEMICAL ANALYSIS; DOSES; MEDICINE; MOCKUP; NONDESTRUCTIVE ANALYSIS; NUCLEAR MEDICINE; RADIOLOGY; SCATTERING; STRUCTURAL MODELS; THERAPY