Published March 2009 | Version v1
Report Open

A high energy hemispherical analyzer for experimental station of PES beamline at Indus-2

  • 1. Technical Physics and Prototype Engineering Division, Bhabha Atomic Research Centre, Mumbai (India)

Description

This article describes the design and development of Hemispherical Deflector Analyzer that is especially designed to analyze high kinetic energy electrons (up to 15 keV) with an energy resolution (ΔE/E) of the order of ∼10-4 for experimental station of PES beamline (BL-14) on Indus-2, a third generation synchrotron source at RRCAT, Indore. With such a high KE analyzing capacity, this analyzer facilitates to measure bulk electronic properties of solids by the Photoemission Spectroscopy with Synchrotron Radiation preserving an energy resolution comparable to modern surface sensitive photoemission spectra. In order to attain better energy resolution over a wider kinetic energy range of electrons, we have designed and constructed a dedicated input lens system for the analyzer. It is further connected with a high stability power supplies, a low dark-count delay line detector and read-out electronics. The system has been designed and fabrication work is almost complete for testing. (author)

Availability note (English)

Available from INIS in electronic form

Files

40051721.pdf

Files (517.2 kB)

Name Size Download all
md5:01f33f610d5f2d7a1f319638c23300ee
517.2 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
19 p.
Report number
BARC--2009/E/006

Optional Information

Notes
31 refs., 4 figs., 2 tabs.