The CDF Run-II silicon microstrip detector has seen almost 12 fb-1 of proton-antiproton collisions over the last 10 years. It has shown remarkable performance, with 80% of its channels still operating error-free, and only one of its eight layers approaching the operational limits for full depletion. The measured depletion voltage and signal-to-noise ratio of these sensors give unique information about the behavior of sensors irradiated slowly over a long period of time. Data from heavily irradiated, double-sided sensors excludes a monotonic electric field inside the sensor and is instead consistent with a doubly-peaked field that is lower in the center of the sensor and higher at the edges.

Availability note (English)

Available from http://lss.fnal.gov/cgi-bin/find_paper.pl?conf-11-447.pdf; PURL: https://www.osti.gov/servlets/purl/1033325/

Additional details

Publishing Information

Imprint Pagination
10 p.
Report number
FERMILAB-CONF--11-447-PPD

Conference

Title
20. Anniversary International Workshop on Vertex Detectors
Dates
19-24 Jun 2011
Place
Rust, Lake Neusiedl (Austria)

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
43017623
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S43: PARTICLE ACCELERATORS;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
AGING; ELECTRIC FIELDS; FERMILAB COLLIDER DETECTOR; IRRADIATION; PERFORMANCE; SENSORS; SI SEMICONDUCTOR DETECTORS; SIGNAL-TO-NOISE RATIO; SILICON
Descriptors DEC
DIMENSIONLESS NUMBERS; ELEMENTS; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SEMIMETALS

Optional Information

Contract/Grant/Project number
AC02-07CH11359
Funding organization
DOE Office of Science (United States)