The fan-beam short-scan filtered backprojection (FBP) algorithm is popular in the computed tomography (CT) industry and in clinical applications. This paper points out that the fan-beam short-scan FBP algorithm is not exact, even in an ideal noiseless situation. The error depends on the focal length of the imaging geometry. When the focal length is long enough, the error is small and can therefore be ignored. (note)
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/60/8/N131Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 60
- Journal Issue
- 8
- Journal Page Range
- p. N131-N139
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47084588
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; BEAMS; BIOMEDICAL RADIOGRAPHY; COMPUTERIZED TOMOGRAPHY; ERRORS
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; MATHEMATICAL LOGIC; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; TOMOGRAPHY