The goal of this prospective study was to compare a full-field digital mammography system (FFDM) to a conventional screen-film mammography system (SFM) for the detection and characterization of microcalcifications. Fifty-five patients with 57 isolated microcalcification clusters were examined using a FFDM system (Senographe 2000D, GE Medical Systems, Milwaukee, Wis.) and a SFM system (Senographe DMR, GE Medical Systems, Milwaukee, Wis.). A conventional screen-film mammogram and a digital contact mammogram were obtained of each cluster. The image quality and the number of calcification particles were evaluated, and a characterization (BI-RADS 1-5) of microcalcifications was given by four experienced readers. Histopathology revealed 16 benign lesions (sclerosing adenosis, dysplasia, hamartoma, radial scar) in 15 patients and 21 malignant tumors (in situ carcinoma, invasive carcinoma) in 20 patients. Twenty patients had benign changes verified by long-term follow-up. Image quality of FFDM was assessed as superior to SFM in more than 50% of the cases. The FFDM showed more calcifications in 41% of all cases. Sensitivity and specificity for FFDM vs SFM were 95.2 vs 91.9% and 41.4 vs 39.3%, respectively. Moreover, FFDM demonstrated a higher diagnostic accuracy (deviation: 0.86 BI-RADS steps) compared with FSM (deviation 0.93 BI-RADS steps). The FFDM system with a 100-μm pixel size provides better image quality than SFM in patients with mammographic microcalcifications. The FFDM has a higher sensitivity and a higher reliability in characterizing microcalcifications. (orig.)

Availability note (English)

Available from: http://dx.doi.org/10.1007/s00330-002-1354-x

Additional details

Identifiers

Publishing Information

Journal Title
European Radiology
Journal Volume
12
Journal Issue
11
Journal Page Range
p. 2679-2683
ISSN
0938-7994
CODEN
EURAE3

INIS