Intraoperative Sac Pressure Measurement During Endovascular Abdominal Aortic Aneurysm Repair
Creators
- 1. Aichi Medical University Hospital, Department of Vascular Surgery (Japan)
- 2. Aichi Medical University Hospital, Department of Radiology (Japan)
Description
PurposeIntraoperative sac pressure was measured during endovascular abdominal aortic aneurysm repair (EVAR) to evaluate the clinical significance of sac pressure measurement.MethodsA microcatheter was placed in an aneurysm sac from the contralateral femoral artery, and sac pressure was measured during EVAR procedures in 47 patients. Aortic blood pressure was measured as a control by a catheter from the left brachial artery.ResultsThe systolic sac pressure index (SPI) was 0.87 ± 0.10 after main-body deployment, 0.63 ± 0.12 after leg deployment (P < 0.01), and 0.56 ± 0.12 after completion of the procedure (P < 0.01). Pulse pressure was 55 ± 21 mmHg, 23 ± 15 mmHg (P < 0.01), and 16 ± 12 mmHg (P < 0.01), respectively. SPI showed no significant differences between the Zenith and Excluder stent grafts (0.56 ± 0.13 vs. 0.54 ± 0.10, NS). Type I endoleak was found in seven patients (15%), and the SPI decreased from 0.62 ± 0.10 to 0.55 ± 0.10 (P = 0.10) after fixing procedures. Type II endoleak was found in 12 patients (26%) by completion angiography. The SPI showed no difference between type II endoleak positive and negative (0.58 ± 0.12 vs. 0.55 ± 0.12, NS). There were no significant differences between the final SPI of abdominal aortic aneurysms in which the diameter decreased in the follow-up and that of abdominal aortic aneurysms in which the diameter did not change (0.53 ± 0.12 vs. 0.57 ± 0.12, NS).ConclusionsSac pressure measurement was useful for instant hemodynamic evaluation of the EVAR procedure, especially in type I endoleaks. However, on the basis of this small study, the SPI cannot be used to reliably predict sac growth or regression.
Additional details
Identifiers
Publishing Information
- Journal Title
- Cardiovascular and Interventional Radiology
- Journal Volume
- 33
- Journal Issue
- 5
- Journal Page Range
- p. 939-942
- ISSN
- 0174-1551
- CODEN
- CAIRDG
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 42028976
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ABDOMEN; AORTA; BASES; BIOMEDICAL RADIOGRAPHY; BLOOD PRESSURE; PRESSURE MEASUREMENT
- Descriptors DEC
- ARTERIES; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY
Optional Information
- Copyright
- Copyright (c) 2010 Springer Science+Business Media, LLC and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE)