Specification and estimation of sources of bias affecting neurological studies in PET/MR with an anatomical brain phantom

  • 1. Turku PET Centre, Turku (Finland)
  • 2. Department of Mathematics and Statistics, University of Turku, Turku (Finland)

Description

Selection of reconstruction parameters has an effect on the image quantification in PET, with an additional contribution from a scanner-specific attenuation correction method. For achieving comparable results in inter- and intra-center comparisons, any existing quantitative differences should be identified and compensated for. In this study, a comparison between PET, PET/CT and PET/MR is performed by using an anatomical brain phantom, to identify and measure the amount of bias caused due to differences in reconstruction and attenuation correction methods especially in PET/MR. Differences were estimated by using visual, qualitative and quantitative analysis. The qualitative analysis consisted of a line profile analysis for measuring the reproduction of anatomical structures and the contribution of the amount of iterations to image contrast. The quantitative analysis consisted of measurement and comparison of 10 anatomical VOIs, where the HRRT was considered as the reference. All scanners reproduced the main anatomical structures of the phantom adequately, although the image contrast on the PET/MR was inferior when using a default clinical brain protocol. Image contrast was improved by increasing the amount of iterations from 2 to 5 while using 33 subsets. Furthermore, a PET/MR-specific bias was detected, which resulted in underestimation of the activity values in anatomical structures closest to the skull, due to the MR-derived attenuation map that ignores the bone. Thus, further improvements for the PET/MR reconstruction and attenuation correction could be achieved by optimization of RAMLA-specific reconstruction parameters and implementation of bone to the attenuation template. -- Highlights: • Comparison between PET, PET/CT and PET/MR was performed with a novel brain phantom. • The performance of reconstruction and attenuation correction in PET/MR was studied. • A recently developed brain phantom was found feasible for PET/MR imaging. • Contrast reduction due to unoptimized RAMLA reconstruction algorithm was detected. • Activity value reduction due to attenuation value assignment in MRAC was detected

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2013.09.002

Additional details

Identifiers

DOI
10.1016/j.nima.2013.09.002;
PII
S0168-9002(13)01236-9;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
734
Journal Issue
Part B
Journal Page Range
p. 179-184
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
2. conference on PET-MR and SPECT-MR: Current status of instrumentation, applications and developments
Acronym
PSMR2013
Dates
6-7 May 2013
Place
Aachen (Germany)

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
45070465
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ATTENUATION; BRAIN; COMPARATIVE EVALUATIONS; IMAGES; PHANTOMS; SKULL
Descriptors DEC
BODY; CENTRAL NERVOUS SYSTEM; EVALUATION; MOCKUP; NERVOUS SYSTEM; ORGANS; SKELETON; STRUCTURAL MODELS

Optional Information

Copyright
Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.