Radon leakage problem in determination of radon by liquid scintillation
Creators
- 1. Institute of Radiation Medicine, Chinese Academy of Medical Sciences (China)
Description
The paper presents briefly the experimental data on radon leakage from liquid scintillation vials in measurement of radon with liquid scintillation counter. Seal tests of vials of three kinds have been made. Results show that all tested caps of the vials made of glass can not meet the needs to seal radon in the vial. The problem was basically solved by putting a layer of rubber (bicycle inner tube, 1 mm thick) and a layer of aluminized polyester (0.01 mm thick) between the vial and the cap as cushion in place of the polyethylene inner cap or other types of gasket. It is suggested that seal ability to radon of vials be tested before using a new brand of commercial scintillation vials for radon determination and remedial action, if necessary, be taken to guarantee the accuracy of analytical results
Additional details
Publishing Information
- Journal Title
- Radiation Protection (Taiyuan)
- Journal Volume
- 12
- Journal Issue
- 4
- Journal Page Range
- p. 322-325.
- ISSN
- 1000-8187
- CODEN
- FUFAEM
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 25029097
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- LEAKS; LIQUID SCINTILLATION DETECTORS; RADON; SCINTILLATION COUNTERS; SEALS
- Descriptors DEC
- ELEMENTS; MEASURING INSTRUMENTS; NONMETALS; RADIATION DETECTORS; RARE GASES