Published December 2003 | Version v1
Journal article

Use of a range scaling method to determine alanine/water stopping power ratios

Description

A phantom composed of alanine dosimeter material has been constructed and depth-dose measurements made in a 10 MeV electron beam. The results have demonstrated the feasibility of using relative depth-dose measurements to determine stopping power ratios in materials of dosimetric interest. Experimental stopping power ratios for alanine dosimeter material and water agreed with the data of ICRU Report 37 within the uncertainty of the experiment (±1.2% at a 95% confidence level)

Additional details

Additional titles

Augmented title (English)
Electron; Stopping power; Absorbed dose; Dosimetry; Alanine

Identifiers

DOI
10.1016/j.radphyschem.2003.10.001;
PII
S0969806X03005267;

Publishing Information

Journal Title
Radiation Physics and Chemistry (1993)
Journal Volume
68
Journal Issue
6
Journal Page Range
p. 1017-1022
ISSN
0969-806X
CODEN
RPCHDM

Optional Information

Copyright
Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.