Published May 15, 1985
| Version v1
Journal article
Scintillation counter for soft X-ray spectroscopy
Creators
- 1. State Univ. of New York, Stony Brook (USA). Dept. of Physics
- 2. Brookhaven National Lab., Upton, NY (USA). Dept. of Chemistry
Description
The phosphor P31 is shown to be a suitable and convenient detector for spectroscopy in the soft X-ray region. Visible light emitted by the phosphor is detected by a photomultiplier tube. The detector is compatible with vacuum, but does not require it. To illustrate its capabilities, gas phase spectra of CO2 in the region of the oxygen K edge are presented. (orig.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods Phys. Res., Sect. A
- Journal Volume
- 236
- Journal Issue
- 2
- Series
- CODEN: NIMRD.;Nucl. Instrum. Methods Phys. Res., Sect. A.
- Journal Page Range
- 419-422
- ISSN
- 0168-9002
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 16072336
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CARBON DIOXIDE; ENERGY SPECTRA; EV RANGE 100-1000; PHOSPHORS; PHOTOMULTIPLIERS; SOFT X RADIATION; SOLID SCINTILLATION DETECTORS; X-RAY DETECTION; X-RAY SPECTRA; X-RAY SPECTROSCOPY
- Descriptors DEC
- CARBON COMPOUNDS; CARBON OXIDES; CHALCOGENIDES; DETECTION; ELECTROMAGNETIC RADIATION; ENERGY RANGE; EV RANGE; IONIZING RADIATIONS; MEASURING INSTRUMENTS; OXIDES; OXYGEN COMPOUNDS; PHOTOTUBES; RADIATION DETECTION; RADIATION DETECTORS; RADIATIONS; SCINTILLATION COUNTERS; SPECTRA; SPECTROSCOPY; X RADIATION
Optional Information
- Contract/Grant/Project number
- Contract DE-AC02-76CH00016