Radiometric and signal-to-noise ratio properties of multiplex dispersive spectrometry
Description
Recent theoretical investigations have shown important radiometric disadvantages of interferential multiplexing in Fourier transform spectrometry that apparently can be applied even to coded aperture spectrometers. We have reexamined the methods of noninterferential multiplexing in order to assess their signal-to-noise ratio (SNR) performance, relying on a theoretical modeling of the multiplexed signals. We are able to show that quite similar SNR and radiometric disadvantages affect multiplex dispersive spectrometry. The effect of noise on spectral estimations is discussed.
Additional details
Identifiers
- DOI
- 10.1364/AO.49.005366;
Publishing Information
- Journal Title
- Applied Optics
- Journal Volume
- 49
- Journal Issue
- 28
- Journal Page Range
- p. 5366-5373
- ISSN
- 0003-6935
- CODEN
- APOPAI
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 43126041
- Subject category
- S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
- Descriptors DEI
- APERTURES; FOURIER TRANSFORM SPECTROMETERS; NOISE; PERFORMANCE; SIGNAL-TO-NOISE RATIO; SIMULATION; SPECTROMETERS; SPECTROSCOPY
- Descriptors DEC
- DIMENSIONLESS NUMBERS; MEASURING INSTRUMENTS; OPENINGS; SPECTROMETERS
Optional Information
- Notes
- (c) 2010 Optical Society of America