A radioimmunoassay of human prealbumin in body fluids
Description
Prealbumin (PA) was purified 35-fold from human serum and antibodies raised against it in rabbits. A 2 hour radioimmunoassay (RIA) using polyethyleneglycol (PEG) to separate bound and free PA was used to determine levels in body fluids. Using patient serum specimens the new method was compared with an electroimmunoassay (EIA) method and the regression equation obtained was: y=1.13x-9.91. The RIA and EIA methods compared favourably with respect to precision to practicability and economy. The RIA method seems especially suitable for large scale assays of PA and is 100 times more sensitive than EIA. Preliminary estimations of PA with the RIA method in plasma, cerebrospinal fluids, amniotic fluids, duodenal juices and urines were carried out. The results indicate that this method can be conveniently used to assay PA in body fluids where the protein is present in low concentration. (Auth.)
Additional details
Publishing Information
- Journal Title
- Clin. Chim. Acta
- Journal Volume
- 134
- Journal Issue
- 1-2
- Series
- Clin. Chim. Acta.
- Journal Page Range
- 69-76
- ISSN
- 0009-8981
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 15013403
- Subject category
- S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- ACCURACY; ALBUMINS; BODY FLUIDS; COMPARATIVE EVALUATIONS; IODINE 125; QUANTITY RATIO; RADIOIMMUNOASSAY; SENSITIVITY
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BIOLOGICAL MATERIALS; DAYS LIVING RADIOISOTOPES; ELECTRON CAPTURE RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; IODINE ISOTOPES; ISOTOPE APPLICATIONS; ISOTOPES; MATERIALS; NUCLEI; ODD-EVEN NUCLEI; ORGANIC COMPOUNDS; PROTEINS; RADIOISOTOPES; TRACER TECHNIQUES