Performance analysis of a new positron camera geometry for high speed, fine particle tracking
- 1. Department of Mining and Materials Engineering, McGill University 3610 University Street, Montreal, Quebec H3A 0C5 (Canada)
- 2. RAP Innovation and Development Pty Ltd, PO Box 559, Indooroopilly, Queensland 4068 (Australia)
- 3. School of Physics and Astronomy, University of Birmingham, Edgbaston, Birmingham B15 2TT (United Kingdom)
Description
A new positron camera arrangement was assembled using 16 ECAT951 modular detector blocks. A closely packed, cross pattern arrangement was selected to produce a highly sensitive cylindrical region for tracking particles with low activities and high speeds. To determine the capabilities of this system a comprehensive analysis of the tracking performance was conducted to determine the 3D location error and location frequency as a function of tracer activity and speed. The 3D error was found to range from 0.54 mm for a stationary particle, consistent for all tracer activities, up to 4.33 mm for a tracer with an activity of 3 MBq and a speed of 4 m · s−1. For lower activity tracers (<10−2 MBq), the error was more sensitive to increases in speed, increasing to 28 mm (at 4 m · s−1), indicating that at these conditions a reliable trajectory is not possible. These results expanded on, but correlated well with, previous literature that only contained location errors for tracer speeds up to 1.5 m · s−1. The camera was also used to track directly activated mineral particles inside a two-inch hydrocyclone and a 142 mm diameter flotation cell. A detailed trajectory, inside the hydrocyclone, of a −212 + 106 µ m (10−1 MBq) quartz particle displayed the expected spiralling motion towards the apex. This was the first time a mineral particle of this size had been successfully traced within a hydrocyclone, however more work is required to develop detailed velocity fields. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6501/aa7dceAdditional details
Identifiers
Publishing Information
- Journal Title
- Measurement Science and Technology
- Journal Volume
- 28
- Journal Issue
- 9
- Journal Page Range
- [14 p.]
- ISSN
- 0957-0233
- CODEN
- MSTCEP
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 49035381
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CYCLONE SEPARATORS; CYLINDRICAL CONFIGURATION; ERRORS; FINE PARTICLES; FLOTATION; GEOMETRY; PARTICLE TRACKS; PERFORMANCE; POSITIONING; POSITRON CAMERAS; POSITRONS; QUARTZ; THREE-DIMENSIONAL CALCULATIONS; TRAJECTORIES; VELOCITY
- Descriptors DEC
- ANTILEPTONS; ANTIMATTER; ANTIPARTICLES; CAMERAS; CONCENTRATORS; CONFIGURATION; ELEMENTARY PARTICLES; EQUIPMENT; FERMIONS; GAMMA CAMERAS; INERTIAL SEPARATORS; LEPTONS; MATHEMATICS; MATTER; MINERALS; OXIDE MINERALS; PARTICLES; SEPARATION EQUIPMENT; SEPARATION PROCESSES