Published September 21, 2017
| Version v1
Journal article
Performance evaluation of MACACO: a multilayer Compton camera
Creators
- 1. Instituto de Física Corpuscular, CSIC/Universitat de València, Valencia (Spain)
Description
Compton imaging devices have been proposed and studied for a wide range of applications. We have developed a Compton camera prototype which can be operated with two or three detector layers based on monolithic lanthanum bromide () crystals coupled to silicon photomultipliers (SiPMs), to be used for proton range verification in hadron therapy. In this work, we present the results obtained with our prototype in laboratory tests with radioactive sources and in simulation studies. Images of a Na and an Y radioactive sources have been successfully reconstructed. The full width half maximum of the reconstructed images is below 4 mm for a Na source at a distance of 5 cm. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6560/aa8070Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 62
- Journal Issue
- 18
- Journal Page Range
- p. 7321-7341
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 51021211
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- EVALUATION; IMAGES; LANTHANUM BROMIDES; PHOTOMULTIPLIERS; RADIATION SOURCES; RADIOTHERAPY; SILICON; SIMULATION; VERIFICATION
- Descriptors DEC
- BROMIDES; BROMINE COMPOUNDS; ELEMENTS; HALIDES; HALOGEN COMPOUNDS; LANTHANUM COMPOUNDS; LANTHANUM HALIDES; MEDICINE; NUCLEAR MEDICINE; PHOTOTUBES; RADIOLOGY; RARE EARTH COMPOUNDS; SEMIMETALS; THERAPY