Published March 1985 | Version v1
Report

Sample preparation techniques for (p, X) spectrometry

  • 1. Department of Scientific and Industrial Research, Lower Hutt (New Zealand). Inst. of Nuclear Sciences

Description

Samples are ashed at low temperature, using oxygen plasma; a rotary evaporator, and freeze drying speeded up the ashing. The new design of apparatus manufactured was only 10 watt but was as efficient as a 200 watt commercial machine; a circuit diagram is included. Samples of hair and biopsy samples of skin were analysed by the technique. A wool standard was prepared for interlaboratory comparison exercises. It was based on New Zealand merino sheep wool and was 2.9 kg in weight. A washing protocol was developed, which preserves most of the trace element content. The wool was ground in liquid nitrogen using a plastic pestle and beaker, driven by a rotary drill press. (author)

Additional details

Publishing Information

Imprint Title
Health-related monitoring of trace element pollutants using nuclear techniques
Imprint Pagination
234 p.
Journal Page Range
p. 211-218.
Report number
IAEA-TECDOC--330

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
16049219
Subject category
S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
Descriptors DEI
CALIBRATION STANDARDS; DRY ASHING; ELEMENTS; HAIR; LABORATORY EQUIPMENT; MAN; PIXE ANALYSIS; SAMPLE PREPARATION; SHEEP; SKIN; TRACE AMOUNTS; WET ASHING; WOOL
Descriptors DEC
ANIMALS; BODY; CHEMICAL ANALYSIS; DOMESTIC ANIMALS; EQUIPMENT; MAMMALS; NONDESTRUCTIVE ANALYSIS; ORGANS; PRIMATES; RUMINANTS; VERTEBRATES; X-RAY EMISSION ANALYSIS