Measurement of leg length discrepancy after total hip arthroplasty. The reliability of a plain radiographic method compared to CT-scanogram
Creators
- 1. Sundsvall Hospital, Department of Orthopaedics, Sundsvall (Sweden)
- 2. Karolinska University Hospital, Huddinge, Department of Radiology, Stockholm (Sweden)
- 3. Vaesternorrland County, Department of Research and Development, Sundsvall (Sweden)
- 4. Umeaa University, Department of Surgical and Perioperative Sciences, Umeaa (Sweden)
Description
To measure the interobserver reliability and intraobserver reproducibility of post total hip arthroplasty (THA) leg length discrepancy (LLD) measurement on radiographs as well as to evaluate its accuracy by comparing it with LLD measurement on computed tomographic scanogram (CT-scanogram). In this prospective study, postoperative LLD measurements in ten THA patients were made by four observers on anteroposterior radiographs of the pelvis (inter-teardrop line to the tip of lesser trochanter) and compared to LLD measurements made on CT-scanogram scout views of the lower limb. Two observers repeated the LLD measurements on radiographs 8 weeks after the first measurements. The interobserver reliability of the LLD measurement on plain radiographs was evaluated by comparing the measurements of the four observers and the intraobserver reproducibility by comparing the two repeated measurements made by the two observers. We found excellent interobserver reliability (mean ICC 0.83) and intraobserver reproducibility (ICC 0.90 and 0.88) of the LLD measurements on plain radiographs. There was a moderate to excellent agreement, but with wide variation of measurements among the four observers, when plain radiographic measurement was compared with CT-scanogram (ICC 0.58, 0.60, 0.71, and 0.82). Despite the excellent interobserver reliability and intraobserver reproducibility of LLD measurement on radiographs, clinicians should be aware of its limited accuracy when compared to CT-scanogram. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00256-011-1166-7Additional details
Identifiers
Publishing Information
- Journal Title
- Skeletal Radiology
- Journal Volume
- 41
- Journal Issue
- 2
- Journal Page Range
- p. 187-191
- ISSN
- 0364-2348
- CODEN
- SKRADI
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 43043876
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; BIOMEDICAL RADIOGRAPHY; BONE JOINTS; COMPUTERIZED TOMOGRAPHY; FEMUR; IMPLANTS; PELVIS; PROSTHESES; RELIABILITY; SENSITIVITY; SPECIFICITY; SURGERY; X-RAY RADIOGRAPHY
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; INDUSTRIAL RADIOGRAPHY; MATERIALS TESTING; MEDICAL SUPPLIES; MEDICINE; NONDESTRUCTIVE TESTING; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; SKELETON; TESTING; TOMOGRAPHY