Characterization measurement of a thick CdTe detector for BNCT-SPECT – Detection efficiency and energy resolution
- 1. Division of Electrical, Electronic and Information Engineering, Graduate School of Engineering, Osaka University, Yamada-oka 2-1, Suita, Osaka 565-0871 (Japan)
- 2. Radiation Research Center, Osaka Prefecture University, Gakuen-cho 1-1, Nakaku, Sakai, Osaka 599-8531 (Japan)
- 3. Department of Oral and Maxillofacial Surgery II, Graduate School of Dentistry, Osaka University, Yamada-oka 1-8, Suita, Osaka 565-0871 (Japan)
Description
Author's group is carrying out development of BNCT-SPECT with CdTe device, which monitors the therapy effect of BNCT in real-time. From the design calculations, the dimensions were fixed to 1.5×2×30 mm3. For the collimator it was confirmed that it would have a good spatial resolution and simultaneously the number of counts would be acceptably large. After producing the CdTe crystal, the characterization measurement was carried out. For the detection efficiency an excellent agreement between calculation and measurement was obtained. Also, the detector has a very good energy resolution so that gamma-rays of 478 keV and 511 keV could be distinguished in the spectrum. - Highlights: • BNCT-SPECT is developed with CdTe device to estimate therapy effect of BNCT. • By design calculations, CdTe dimensions are determined to be 1.5×2×30 mm3. Collimator length is 10 cm with 2 mm diameter hole. • Producing the crystal, efficiency and energy resolution were measured. • Excellent agreement was obtained between measurement and calculation. Discrimination of 478 keV and 511 keV was confirmed in the spectrum
Availability note (English)
Available from http://dx.doi.org/10.1016/j.apradiso.2014.01.023Additional details
Identifiers
- DOI
- 10.1016/j.apradiso.2014.01.023;
- PII
- S0969-8043(14)00036-0;
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 88
- Journal Issue
- Complete
- Journal Page Range
- p. 129-133
- ISSN
- 0969-8043
- CODEN
- ARISEF
Conference
- Title
- 15. international congress on neutron capture therapy impact of a new radiotherapy against cancer
- Dates
- 23 Aug 2005
- Place
- Tsukuba (Japan)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46102928
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BORON 10; CDTE SEMICONDUCTOR DETECTORS; COLLIMATORS; CRYSTALS; DESIGN; EFFICIENCY; ENERGY RESOLUTION; GAMMA RADIATION; KEV RANGE; LITHIUM 7; MONITORS; NEUTRON CAPTURE THERAPY; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY; SPATIAL RESOLUTION
- Descriptors DEC
- BORON ISOTOPES; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; EMISSION COMPUTED TOMOGRAPHY; ENERGY RANGE; IONIZING RADIATIONS; ISOTOPES; LIGHT NUCLEI; LITHIUM ISOTOPES; MEASURING INSTRUMENTS; MEDICINE; NEUTRON THERAPY; NUCLEAR MEDICINE; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; RADIATION DETECTORS; RADIATIONS; RADIOLOGY; RADIOTHERAPY; RESOLUTION; SEMICONDUCTOR DETECTORS; STABLE ISOTOPES; THERAPY; TOMOGRAPHY
Optional Information
- Copyright
- Copyright (c) 2014 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.