Published 1975 | Version v1
Report Open

U isotope calibration standards

  • 1. CEA Centre de Pierrelatte, 26 (France)

Description

A new technique for the preparation of U isotope calibration standards is described. The principle is the following: two solutions with the same concentration in U are prepared, one being highly enriched in 235U and the other in 238U. A calibration standard of intermediate isotope composition is prepared by mixing an equal weight of these solutions. By successive dilution, it is possible to obtain a series of primary calibration standards

Abstract (French)

On decrit une nouvelle technique de preparation des etalons isotopiques d'uranium. Le principe est le suivant: deux solutions identiques en concentration d'uranium sont preparees, l'une tres riche en 235U et l'autre en 238U. En melangeant par poids egaux ces solutions, on realise un etalon de composition isotopique intermediaire. Par dilutions isotopiques successives, on peut preparer une serie d'etalons primaires

Files

8346095.pdf

Files (143.0 kB)

Name Size Download all
md5:fd13fd8e35bec97d007d980f5f85cd84
143.0 kB Preview Download

System files (11.5 kB)

Name Size Download all

Additional details

Additional titles

Original title (French)
Etalons isotopiques d'uranium

Publishing Information

Imprint Title
Meeting on uranium analysis. Saclay, 17-18 December 1975
Journal Page Range
p. 149-154.
Report number
CEA-CR--11

Conference

Title
Meeting on uranium analysis.
Dates
17 - 18 Dec 1975.
Place
Saclay, France.

INIS

Optional Information

Notes
Imprint:Reunion technique sur les methodes d'analyse de l'uranium. Saclay, 17-18 decembre 1975.