Precision-controlled elution of a 82Sr/82Rb generator for cardiac perfusion imaging with positron emission tomography
Creators
- 1. National Cardiac PET Centre, University of Ottawa Heart Institute, Ottawa, Ontario (Canada)
- 2. School of Information Technology and Engineering, University of Ottawa, Ottawa, Ontario (Canada)
Description
A rubidium-82 (82Rb) elution system is described for use with positron emission tomography. Due to the short half-life of 82Rb (76 s), the system physics must be modelled precisely to account for transport delay and the associated activity decay and dispersion. Saline flow is switched between a 82Sr/82Rb generator and a bypass line to achieve a constant-activity elution of 82Rb. Pulse width modulation (PWM) of a solenoid valve is compared to simple threshold control as a means to simulate a proportional valve. A predictive-corrective control (PCC) algorithm is developed which produces a constant-activity elution within the constraints of long feedback delay and short elution time. The system model parameters are adjusted through a self-tuning algorithm to minimize error versus the requested time-activity profile. The system is self-calibrating with 2.5% repeatability, independent of generator activity and elution flow rate. Accurate 30 s constant-activity elutions of 10-70% of the total generator activity are achieved using both control methods. The combined PWM-PCC method provides significant improvement in precision and accuracy of the requested elution profiles. The 82Rb elution system produces accurate and reproducible constant-activity elution profiles of 82Rb activity, independent of parent 82Sr activity in the generator. More reproducible elution profiles may improve the quality of clinical and research PET perfusion studies using 82Rb
Additional details
Identifiers
- DOI
- 10.1088/0031-9155/52/3/009;
- PII
- S0031-9155(07)28986-9;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 52
- Journal Issue
- 3
- Journal Page Range
- p. 659-673
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38072771
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; ALGORITHMS; BYPASSES; FLOW RATE; MODULATION; POSITRON COMPUTED TOMOGRAPHY; RUBIDIUM 82; SOLENOIDS; STRONTIUM 82
- Descriptors DEC
- ALKALINE EARTH ISOTOPES; BETA DECAY RADIOISOTOPES; BETA-PLUS DECAY RADIOISOTOPES; COMPUTERIZED TOMOGRAPHY; DAYS LIVING RADIOISOTOPES; DIAGNOSTIC TECHNIQUES; ELECTRIC COILS; ELECTRICAL EQUIPMENT; ELECTRON CAPTURE RADIOISOTOPES; EMISSION COMPUTED TOMOGRAPHY; EQUIPMENT; EVEN-EVEN NUCLEI; HOURS LIVING RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; ISOTOPES; MATHEMATICAL LOGIC; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-ODD NUCLEI; RADIOISOTOPES; RUBIDIUM ISOTOPES; STRONTIUM ISOTOPES; TOMOGRAPHY