PET functional volume delineation: a robustness and repeatability study
Creators
- 1. CHU MORVAN, LaTIM, INSERM U650, Brest (France)
- 2. CHU Morvan, INSERM, U650, LaTIM, Brest (France)
- 3. CHU, Academic Department of Nuclear Medicine, Brest (France)
- 4. CHU, Institute of Oncology, Brest (France)
Description
Current state-of-the-art algorithms for functional uptake volume segmentation in PET imaging consist of threshold-based approaches, whose parameters often require specific optimization for a given scanner and associated reconstruction algorithms. Different advanced image segmentation approaches previously proposed and extensively validated, such as among others fuzzy C-means (FCM) clustering, or fuzzy locally adaptive bayesian (FLAB) algorithm have the potential to improve the robustness of functional uptake volume measurements. The objective of this study was to investigate robustness and repeatability with respect to various scanner models, reconstruction algorithms and acquisition conditions. Robustness was evaluated using a series of IEC phantom acquisitions carried out on different PET/CT scanners (Philips Gemini and Gemini Time-of-Flight, Siemens Biograph and GE Discovery LS) with their associated reconstruction algorithms (RAMLA, TF MLEM, OSEM). A range of acquisition parameters (contrast, duration) and reconstruction parameters (voxel size) were considered for each scanner model, and the repeatability of each method was evaluated on simulated and clinical tumours and compared to manual delineation. For all the scanner models, acquisition parameters and reconstruction algorithms considered, the FLAB algorithm demonstrated higher robustness in delineation of the spheres with low mean errors (10%) and variability (5%), with respect to threshold-based methodologies and FCM. The repeatability provided by all segmentation algorithms considered was very high with a negligible variability of <5% in comparison to that associated with manual delineation (5-35%). The use of advanced image segmentation algorithms may not only allow high accuracy as previously demonstrated, but also provide a robust and repeatable tool to aid physicians as an initial guess in determining functional volumes in PET. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00259-010-1688-6Additional details
Identifiers
Publishing Information
- Journal Title
- European Journal of Nuclear Medicine and Molecular Imaging
- Journal Volume
- 38
- Journal Issue
- 4
- Journal Page Range
- p. 663-672
- ISSN
- 1619-7070
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 43018606
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; ALGORITHMS; CARCINOMAS; CLASSIFICATION; COMPUTERIZED TOMOGRAPHY; DATA ACQUISITION; ERRORS; FUZZY LOGIC; IMAGE PROCESSING; PHANTOMS; POSITRON COMPUTED TOMOGRAPHY; VALIDATION; VOLUME
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DISEASES; EMISSION COMPUTED TOMOGRAPHY; MATHEMATICAL LOGIC; MOCKUP; NEOPLASMS; PROCESSING; STRUCTURAL MODELS; TESTING; TOMOGRAPHY