Published 2003 | Version v1
Miscellaneous

BH1216III two-sample low-background α and β scintillation counter

  • 1. Beijing Nuclear Instruments Factory, Beijing (China)

Description

This report describes an α and β radiation detection system capable of counting emissions from two samples simultaneously. Thin plastic phosphors and zinc sulphide phosphors are used as primary detectors, shielded against background radiation by a single plastic scintillator guard detector coupled in anticoincidence with the primary detectors. 239Pu α counting efficiency is 72% with a background of 1.59 counts per hour. 90Sr-90Y β counting efficiency is 45% with a background of 1.11 counts per minute

Part of:
Proceedings of the 11th national conference on nuclear electronics and nuclear detection technology

Additional details

Publishing Information

Imprint Title
Proceedings of the 11th national conference on nuclear electronics and nuclear detection technology
Imprint Pagination
598 p.
Journal Page Range
p. 288-293

Conference

Title
11. national conference on nuclear electronics and nuclear detection technology
Dates
1-6 Dec 2002
Place
Xiamen (China)

Optional Information