Comparison of conventional hand examination on six optimised DR systems
Creators
- 1. Department of Regional Health Research, University of Southern Denmark, J.B. Winslows Vej 19, 3., 5000 Odense C (Denmark)
- 2. Department of Radiology, Kolding, Lillebaelt Hospital, University Hospitals of Southern Denmark, Sygehusvej 24, 6000 Kolding (Denmark)
- 3. Health Sciences Research Centre, UCL University College, Niels Bohrs Alle 1, 5230 Odense M (Denmark)
- 4. Department of Radiology and Nuclear Medicine, University Hospital of Southwest Denmark, Finsensgade 35, 6700 Esbjerg (Denmark)
- 5. Department of Health Sciences, Gjovik, Norwegian University of Science and Technology, Teknologiveien 22, 2803 Gjovik (Norway)
Description
The purpose of this study was to investigate the challenges in comparing digital radiography (DR) systems from different vendors for various combinations of exposure factors in posterior-anterior hand radiographs. Image quality was evaluated for a range of tube voltages and tube current-time products using a technical contrast-detail (CDRAD) phantom and an anthropomorphic hand phantom. 900 technical CDRAD images were analysed providing quality figures of merit (IQFinv) and two experienced reporting radiographers using visual grading analysis (VGA) scored 108 anthropomorphic images. This study demonstrates the differences between the DR systems included. When compensating for variations in dose, Canon showed superior results for technical image quality and Fuji for visual image quality for a standard dose point at DR hand examination (ln(DAP) 1.1, 50 kV and 2.5 mAs). (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncab067Additional details
Identifiers
- DOI
- 10.1093/rpd/ncab067;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 194
- Journal Issue
- 1
- Journal Page Range
- p. 27-35
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 52067971
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; IMAGES; PHANTOMS; RADIATION DOSES; RADIOLOGICAL PERSONNEL
- Descriptors DEC
- DOSES; EVALUATION; MEDICAL PERSONNEL; MOCKUP; PERSONNEL; STRUCTURAL MODELS
Optional Information
- Notes
- 25 refs.