A 30 ps Timing Resolution for Single Photons with Multi-pixel Burle MCP-PMT

Description

We have achieved ∼30 psec single-photoelectron and ∼12ps for multi-photoelectron timing resolution with a new 64 pixel Burle MCP-PMT with 10 micron microchannel holes. We have also demonstrated that this detector works in a magnetic field of 15kG, and achieved a single-photoelectron timing resolution of better than 60 psec. The study is relevant for a new focusing DIRC RICH detector for particle identification at future Colliders such as the super B-factory or ILC, and for future TOF techniques. This study shows that a highly pixilated MCP-PMT can deliver excellent timing resolution

Availability note (English)

Available from http://www.slac.stanford.edu/cgi-wrap/getdoc/SLAC-PUB--11934.pdf; http://www.slac.stanford.edu/cgi-wrap/pubpage?SLAC-PUB--11934.html; PURL: https://www.osti.gov/servlets/purl/885512-sR9ern/

Additional details

Publishing Information

Imprint Pagination
4 p.
Report number
SLAC-PUB--11934

Conference

Title
Frontier Detectors for Frontier Physics
Acronym
10. Pisa Meeting on Advanced Detectors
Dates
21-27 May 2006
Place
La Biodola, Elba (Italy)

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
37095671
Subject category
S72: PHYSICS OF ELEMENTARY PARTICLES AND FIELDS;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
FOCUSING; MAGNETIC FIELDS; PARTICLE IDENTIFICATION; PHOTONS; PHYSICS; RESOLUTION
Descriptors DEC
BOSONS; ELEMENTARY PARTICLES; MASSLESS PARTICLES

Optional Information

Contract/Grant/Project number
AC02-76SF00515
Funding organization
US Department of Energy (United States)