Implementation of a program for quality assurance on leaf positioning accuracy using Gafchromic® RTQA2 films

  • 1. Department of Radiation Oncology, School of Specialization in Medical Physics, University of Turin, Turin (Italy)
  • 2. Department of Radiology, University of Turin, Turin (Italy)
  • 3. Radiation Oncology Unit, University of Turin, S. Luigi Gonzaga Hospital, Orbassano (Italy)

Description

In radiotherapy treatments the correct dose delivery to the target volume and the consequent conservation of healthy tissues is affected by multileaf collimator (MLC) leaf positioning accuracy and reproducibility, mostly in intensity-modulated radiation therapy (IMRT): For this reason a quality assurance (QA) program is necessary to ensure the best treatment possible to each patient. The aim of this study is the implementation of a method using Gafchromic® RTQA 2 films to perform routine QA on the MLC, both for qualitative and quantitative analysis. A flatbed document scanner (Epson 10000XL) was used in conjunction with radiochromic detector; a scanning protocol was firstly defined to improve readout accuracy. RTQA2 films were irradiated with 6 MV X-rays at different dose levels to obtain calibration curve. To evaluate the leaf positioning accuracy in different conditions, a rhomboidal shape and a field consisting in three rectangular segments were selected. The images quantitative analysis was handled with a program developed in MATLAB to evaluate the differences between expected and measured leaves positions. The reproducibility and global uncertainty of the method were estimated to be equal to 0.5% and 0.6 mm, respectively. Moreover, a qualitative test was performed: A garden picket fence field, consisting in multiple segments 2 x 22 cm2, was realized setting known leaves shifts to test the method sensitivity. The picket fence test shows that the method is able to detect displacements equal to 1 mm. The results suggest that Gafchromic® RTQA2 films represent a reliable tool to perform MLC routine QA. (author)

Additional details

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
39
Journal Issue
2
Journal Page Range
p. 116-120
CODEN
JMPHFE

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
45078767
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COLLIMATORS; QUALITY ASSURANCE; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY
Descriptors DEC
DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY

Optional Information

Notes
11 refs., 4 figs.