Calibration methods for ECE systems with microwave sources
Description
The authors investigated the feasibility of two methods for calibration of electron cyclotron emission (ECE) systems, both based on the use of a microwave source. In the first method -called the Antenna Pattern Integration (API) method - the microwave source is scanned in space, so as to simulate a large - area - blackbody -source. In the second method -called the Untuned Cavity (UC) method -an untuned cavity, fed by the microwave source, is used to simulate a blackbody. For both methods, the hardware required to perform partly automated calibrations was developed. The microwave based methods were compared with a large area blackbody calibration on two different ECE systems, a Michelson interferometer and a grating polychromator. The API method was found to be more successful than the UC method. (author)