Published June 26, 2013 | Version v1
Journal article

Reconstructing 3D x-ray CT images of polymer gel dosimeters using the zero-scan method

  • 1. School of Chemistry, Physics and Mechanical Engineering, Queensland University of Technology, GPO Box2434, Brisbane, Qld 4001 (Australia)
  • 2. Premion Cancer Care, The Wesley Medical Centre, Suite 1, 40 Chasely St, Auchenflower, Qld 4066 (Australia)

Description

In this study x-ray CT has been used to produce a 3D image of an irradiated PAGAT gel sample, with noise-reduction achieved using the 'zero-scan' method. The gel was repeatedly CT scanned and a linear fit to the varying Hounsfield unit of each pixel in the 3D volume was evaluated across the repeated scans, allowing a zero-scan extrapolation of the image to be obtained. To minimise heating of the CT scanner's x-ray tube, this study used a large slice thickness (1 cm), to provide image slices across the irradiated region of the gel, and a relatively small number of CT scans (63), to extrapolate the zero-scan image. The resulting set of transverse images shows reduced noise compared to images from the initial CT scan of the gel, without being degraded by the additional radiation dose delivered to the gel during the repeated scanning. The full, 3D image of the gel has a low spatial resolution in the longitudinal direction, due to the selected scan parameters. Nonetheless, important features of the dose distribution are apparent in the 3D x-ray CT scan of the gel. The results of this study demonstrate that the zero-scan extrapolation method can be applied to the reconstruction of multiple x-ray CT slices, to provide useful 2D and 3D images of irradiated dosimetry gels.

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/444/1/012091

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
444
Journal Issue
1
Journal Page Range
[4 p.]
ISSN
1742-6596

Conference

Title
7. international conference on 3D radiation dosimetry
Acronym
IC3DDose
Dates
4-8 Dec 2012
Place
Sydney (Australia)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
44074808
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CAT SCANNING; HEATING; IMAGES; IRRADIATION; POLYMER GEL DOSEMETERS; POLYMER GEL DOSIMETRY; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; X RADIATION; X-RAY TUBES
Descriptors DEC
CHEMICAL DOSEMETERS; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DOSEMETERS; DOSES; DOSIMETRY; ELECTROMAGNETIC RADIATION; ELECTRON TUBES; EQUIPMENT; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATIONS; TOMOGRAPHY; X-RAY EQUIPMENT