Published 2006 | Version v1
Journal article

Influence of environmental changes on continuous radon monitors. Results of a Spanish intercomparison exercise

  • 1. Institut de Tecniques Energetiques, Universitat Politecnica de Catalunya, Avda. Diagonal 647, 08028 Barcelona (Spain)

Description

The first Spanish intercomparison exercise for continuous radon monitors was carried out with the participation of nine monitoring systems from eight laboratories. The exposures were carried out in the radon and thoron chambers at the Inst. of Energy Techniques (INTE) of the Technical Univ. of Catalonia (UPC), which is considered to be the Spanish reference chamber. The monitors were exposed to three different temperatures (13, 20 and 30 deg. C), relative humidities (30, 45 and 80%) and radon concentrations (450, 2000 and 9000 Bq m-3). Exposures in the thoron chamber were carried out at concentrations of ∼450 Bq m-3. The response of the ionisation chambers and scintillation monitors was acceptable. However, the response of monitors based on electrostatic collection was found to be influenced by external climatic conditions. Moreover, all radon monitors were sensitive to thoron concentration, which was especially significant for scintillation monitors. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncl036

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
121
Journal Issue
3
Journal Page Range
p. 303-309
ISSN
0144-8420

Optional Information

Notes
8 refs