Long term quality measurements during clinical operation of Limex low energy medical accelerator
- 1. M. Sklodowska-Curie Memorial Cancer Center, Warsaw (Poland). Inst. of Oncology
- 2. Soltan Inst. for Nuclear Studies, Otwock-Swierk (Poland)
- 3. Institute of Nuclear Chemistry and Technology, Warsaw (Poland)
Description
The 4 MeV X-ray beam of Limex accelerator has been applied at Warsaw Cancer Centre since 1986. Prior to clinical use, measurements of beam distributions, including beam quality and output dose, were performed. Its results were compared with the beam distributions of Theratron 780C Cobalt-60 unit. Since 1986, periodical quality measurements have been continuously conducted. Five years experience of Limex exploitation indicates that the dose distributions have been time-stable and agreeable for radiotherapy. (author) 6 refs.; 5 figs.; 1 tab
Additional details
Publishing Information
- Publisher
- Editions Frontieres.
- Imprint Place
- Paris (France)
- ISBN
- 2-86332-115-3
- Imprint Title
- Proceedings of the 3. European particle accelerator conference. V. 2
- Imprint Pagination
- [941 p.].
- Journal Page Range
- p. 1674-1676.
Conference
- Title
- 3. European particle accelerator conference.
- Acronym
- EPAC 92
- Dates
- 24-28 Mar 1992.
- Place
- Berlin (Germany).
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 24064180
- Subject category
- S43: PARTICLE ACCELERATORS; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COMPARATIVE EVALUATIONS; DOSIMETRY; LINEAR ACCELERATORS; NUCLEAR MEDICINE; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY; X-RAY SOURCES
- Descriptors DEC
- ACCELERATORS; EVALUATION; MEDICINE; RADIATION SOURCES; THERAPY