Published February 1989 | Version v1
Journal article

Constant fraction timing with scintillation detectors

  • 1. Faculty of Science, Alexandria Univ., Alexandria (Egypt)

Description

A model is presented for constant fraction pick-off timing with scintillator-photomultiplier detectors based on a statistical method for leading edge timing. Many of the essential features of this technique are obtained such as prompt time-response, the dependence of FWHM on dynamic range of pulse heights and on maximum energy deposited in scintillator; also the effect of delay time on the optimum resolution in CFPHT and ARC timing. The Walk component in this technique is also satisfactory reproduced

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
36
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
401-406
ISSN
0018-9499
CODEN
IETNA

Conference

Title
IEEE nuclear science symposium.
Dates
9-11 Nov 1988.
Place
Orlando, FL (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
20049634
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ALGORITHMS; PERFORMANCE TESTING; PHOTOMULTIPLIERS; SCINTILLATION COUNTERS
Descriptors DEC
MEASURING INSTRUMENTS; PHOTOTUBES; RADIATION DETECTORS; TESTING

Optional Information

Secondary number(s)
CONF-881103--.