Description

I.N.S. type 20 C.P.H.A. is an improved type of and is superior to, Johnstone type P.H.A. in the pulse characteristics. In this paper, its performance and adjustment method are reported. This P.H.A. is suitable for use with 2μs, delay-line shaped pulses having a rise time in the range 0.2 to 0.5μs, a decay time 8μs or more, and a rate 10,000 c/s or less. Adjustment accuracy of channel width is kept in ± 2% and level drift is 0.1 to 0.2 V order. (author)