Published 1987 | Version v1
Report Open

Moisture insensitive charcoal canisters

Description

Continuous monitoring of 222Rn concentrations in the air in houses is the most appropriate approach for the real-time measurements, but this requires complex and expensive instruments and is not practical for large studies. Activated carbon canisters have been used extensively for determining the average concentration over a period of a few days. The ''open face'' charcoal detectors have an integration time constant of about 14 h so that they are sensitive to short-term transient changes in the radon concentration. In addition, water uptake at high relative humidities reduces the radon uptake by the charcoal. The addition of a diffusion barrier and a nylon screen results in a charcoal detector with an integration half-time ranging from 20 to 60 h and a reduced uptake of water at high humidities. Silicone rubber sheeting is relatively permeable to radon and impermeable to water vapor. It was the purpose of this study to evaluate the effect of a silicone barrier on the charcoal canister radon collective device. 3 refs

Availability note (English)

Available from NTIS, PC A02/MF A01 as DE88005916.

Files

19093655.pdf

Files (64.9 kB)

Name Size Download all
md5:80eb091f666fa71e933cf0de88ae38aa
64.9 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
3 p.
Report number
CONF-8709187--1

Conference

Title
Technical exchange meeting on passive radon monitoring.
Dates
21-22 Sep 1987.
Place
Grand Junction, CO (USA).