Published March 1986
| Version v1
Journal article
Moessbauer spectra measuring at high counting rates
Description
High counting rate of γ-quanta when measuring Moessbauer spectra promotes increase of statistical accuracy and reduction of measurement time. At the same time it should be taken into account that errors of electron equipment conditioned by finite dead time of the recording device can result at high counintg rates to errors when measuring spectrum characteristics. Effect of counting lowes of the recorder on characteristics of Moessbauer spectra is studied theoretically and experimentally
Additional details
Additional titles
- Original title (Russian)
- Izmerenie messbauehrovskikh spektrov pri vysokoj skorosti scheta
Publishing Information
- Journal Title
- Prib. Tekh. Ehksp.
- Journal Issue
- no.2
- Series
- Prib. Tekh. Ehksp.
- ISSN
- 0032-8162
- CODEN
- PRTEA
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- USSR
- INIS RN
- 18006378
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ABSORPTION SPECTRA; COUNTING RATES; DEAD TIME; ERRORS; MOESSBAUER SPECTROMETERS; MULTI-CHANNEL ANALYZERS
- Descriptors DEC
- ELECTRONIC EQUIPMENT; EQUIPMENT; GAMMA SPECTROMETERS; MEASURING INSTRUMENTS; PULSE ANALYZERS; SPECTRA; SPECTROMETERS; TIMING PROPERTIES
Optional Information
- Notes
- For English translation see the journal Instruments and Experimental Techniques (USA).