Published March 1986 | Version v1
Journal article

Moessbauer spectra measuring at high counting rates

  • 1. AN SSSR, Novosibirsk. Inst. Neorganicheskoj Khimii

Description

High counting rate of γ-quanta when measuring Moessbauer spectra promotes increase of statistical accuracy and reduction of measurement time. At the same time it should be taken into account that errors of electron equipment conditioned by finite dead time of the recording device can result at high counintg rates to errors when measuring spectrum characteristics. Effect of counting lowes of the recorder on characteristics of Moessbauer spectra is studied theoretically and experimentally

Additional details

Additional titles

Original title (Russian)
Izmerenie messbauehrovskikh spektrov pri vysokoj skorosti scheta

Publishing Information

Journal Title
Prib. Tekh. Ehksp.
Journal Issue
no.2
Series
Prib. Tekh. Ehksp.
ISSN
0032-8162
CODEN
PRTEA

Optional Information

Notes
For English translation see the journal Instruments and Experimental Techniques (USA).