Characterisation of 32P beta source for external radiation protection
Creators
- 1. Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, Mumbai 400094 (India)
- 2. Isotope Production and Applications Division, Bhabha Atomic Research Centre, Mumbai 400094 (India)
Description
Custom-made design of 32P patch sources are used in brachytherapy applications. Use of this source requires external radiation protection. This implies measurement of absorbed dose rate to tissue at a depth of 0.07 mm. Towards this goal, 2.75 mCi 32P patch source was locally developed. Characterization of this source has been carried out by measuring source parameters such as uniformity in the activity distribution and beam uniformity. The characterization also includes output measurements in terms of absorbed dose rate to air using an extrapolation chamber at 11, 20 and 30 cm from the source. Various correction factors which are dependent on the cavity length as well as independent of the cavity length were also determined in this study. The backscatter correction and Bragg–Grapy stopping power ratio were calculated using the Monte Carlo techniques. Using these correction factors and the measured absorbed dose rate to air, absorbed dose rate to tissue at a depth of 0.07 mm was determined. The study also includes measurements of Hp(0.07) and Hp(10) at different source-to-detector distances using an electronic pocket dosimeter. - Highlights: • 32P patch source for external radiation protection. • Characterization of source parameters such as activity and beam uniformity. • Measurement of absorbed dose to tissue using extrapolation chamber. • Determination backscatter factor and stopping power ratio using Monte Carlo technique. • Measurement of Hp(10) and Hp(0.07) using electronic pocket dosimeter.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radmeas.2016.02.034Additional details
Identifiers
- DOI
- 10.1016/j.radmeas.2016.02.034;
- PII
- S1350-4487(16)30069-5;
Publishing Information
- Journal Title
- Radiation Measurements
- Journal Volume
- 89
- Journal Page Range
- p. 52-62
- ISSN
- 1350-4487
- CODEN
- RMEAEP
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48099852
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ABSORBED RADIATION DOSES; ANIMAL TISSUES; BETA SOURCES; DOSE RATES; EXTRAPOLATION CHAMBERS; MONTE CARLO METHOD; PHOSPHORUS 32; RADIATION PROTECTION; STOPPING POWER
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BODY; CALCULATION METHODS; DAYS LIVING RADIOISOTOPES; DOSEMETERS; DOSES; IONIZATION CHAMBERS; ISOTOPES; LIGHT NUCLEI; MEASURING INSTRUMENTS; NUCLEI; ODD-ODD NUCLEI; PARTICLE SOURCES; PHOSPHORUS ISOTOPES; RADIATION DETECTORS; RADIATION DOSES; RADIATION SOURCES; RADIOISOTOPES
Optional Information