On the intrinsic resolution of a gamma camera system
Description
Theoretical simulation of the Anger gamma camera system on a digital computer was used to study the dependence of the intrinsic resolution on the stochastic nature of the photomultiplier pulse formation, as well as on some design parameters. A simple analytical model was found to give similar results on a relative basis. A substantial improvement in resolution is obtained by a proper choice of light-guide material and by introducing specular reflection at the crystal front surface.
Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine & Biology
- Journal Volume
- 17
- Journal Issue
- 4
- Series
- Phys. Med. Biol.
- Journal Page Range
- 514-524
- ISSN
- 0031-9155
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 4046474
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANALYTICAL SOLUTION; DIGITAL SYSTEMS; GAMMA DETECTION; MATHEMATICAL MODELS; PHOTOMULTIPLIERS; PULSE TECHNIQUES; REFLECTION; REFRACTION; RESOLUTION; STOCHASTIC PROCESSES
- Descriptors DEC
- PHOTOTUBES; RADIATION DETECTION