Published 2010 | Version v1
Miscellaneous Open

Response comparative study of Rn-222 alpha particles track monitors

Description

This work was a comparative study between the responses of three monitors, the NRPB, an acrylic monitor (in thin film geometry) and the aluminum monitor (also thin film geometry) in controlled and mixed environment. The experiments consisted on placing the monitors in a plastic tube, with a radio-226 source internal. Only internal CR-39 plastic detectors were analyzed in this work. It was found that the monitors in thin film geometry had activities response of approximately 15% less than the NRPB monitors. All monitors responded the same way when in controlled environment. Related to the type of material, conductive plastic or dielectric (insulator) plastic, the NRPB, in environments without ventilation, responded in the same way. (author)

Files

41123553.pdf

Files (6.5 MB)

Name Size Download all
md5:6ae2b987d494a6587014b4c55b607120
6.5 MB Preview Download

Additional details

Additional titles

Original title (Portuguese)
Estudo comparativo das respostas de monitores de Rn-222 que utilizam o CR-39 como detector de tracos de particulas alfa

Publishing Information

Imprint Pagination
127 p.
Report number
INIS-BR--7699