Published June 2007 | Version v1
Report Open

Study on a new calibration methods of in-situ HPGe γ spectrometers used for non-destructive analyzing radioactivity in nuclear facilities decommissioning

Description

A new calibration technique, which is the Monte Carlo modeling technique, of in-situ HPGe γ spectrometers used for non-destructive analyzing radioactivity in nuclear facilities decommissioning, is presented. A series of assay for some stainless steel pipes and tanks in some nuclear facilities/laboratories of CIAE are taken on site with the in-situ HPGe γ spectrometer. At the same time, some examples are taken and analyzed in laboratories. The relative bias/variation between the values of activity measured by in-situ HPGe γ spectrometers on site and that analyzed in laboratory is less than ±45.0%. (authors)

Availability note (English)

Available from INIS in electronic form

Files

39113940.pdf

Files (709.7 kB)

Name Size Download all
md5:45a921b39f16ef3876c8f4f8c7364af1
709.7 kB Preview Download

Additional details

Publishing Information

ISBN
978-7-5022-3904-6
Imprint Pagination
18 p.
Report number
CNIC--01893

Optional Information

Notes
10 figs., 7 tabs., 5 refs.
Secondary number(s)
IAE--0221