Published May 1988
| Version v1
Journal article
A double grid position sensitive avalanche chamber
Creators
- 1. Academia Sinica, Lanzhou, GS (China). Inst. of Modern Physics
Description
A double grid position sensitive avalanche chamber (DGAC) has been developed. Its sensitive area is 250 x 200 mm2. The properties of the DGAC were tested in flowing isobutylene gas at a pressure of 360 Pa. The position resolution of 4 mm in both x and y coordinates and the time resolution of 290 ps are obtained for 252Cf fission fragment source
Additional details
Publishing Information
- Journal Title
- Nuclear Electronics and Detection Technology
- Journal Volume
- 8
- Journal Issue
- 3
- Series
- Nucl. Electron. Detect. Technol.
- Journal Page Range
- 145-150
- ISSN
- 0258-0934
- CODEN
- HDYUE
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 20042478
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIFORNIUM 252; ELECTRIC FIELDS; ENERGY LOSSES; FABRICATION; FISSION FRAGMENT DETECTION; GRIDS; PERFORMANCE; POSITION SENSITIVE DETECTORS; TIME RESOLUTION; TOWNSEND DISCHARGE
- Descriptors DEC
- ACTINIDE NUCLEI; ALPHA DECAY RADIOISOTOPES; CALIFORNIUM ISOTOPES; DETECTION; ELECTRIC DISCHARGES; ELECTRODES; EVEN-EVEN NUCLEI; HEAVY NUCLEI; ISOTOPES; MEASURING INSTRUMENTS; NUCLEI; RADIATION DETECTION; RADIATION DETECTORS; RADIOISOTOPES; RESOLUTION; SPONTANEOUS FISSION RADIOISOTO; TIMING PROPERTIES; YEARS LIVING RADIOISOTOPES