Published 1999 | Version v1
Journal article

TLD Audit in Radiotherapy in the Czech Republic

  • 1. Prague (Czech Republic)

Description

The National Radiation Protection Institute in Prague organises the TLD audit. The aim of the audit is to provide a basis control of clinical dosimetry in the Czech Republic for purposes of state supervision in radiotherapy. The TLD audit covers absorbed dose measurements under reference conditions for 60Co and 137Cs beams, high energy X ray and electron beams of linear accelerators nd betatrons. Absorbed dose measured by TLD is compared with absorbed dose stated by the radiotherapy centre. Encapsulated LiF:Mg,Ti powder is used for the measurement. A deviation of ±3% between the stated and TLD measured dose is considered acceptable. The first TLD audit was in 1997. A total of 110 beams were checked. Seven major deviations (more than ±6%) were found, which were investigated. Medical physicists from these departments reported a set-up error. However, at most of those hospitals with major deviation, an in situ audit in detail was carried out soon after the TLD audit. Discrepancies were found in clinical dosimetry but some of the irradiation units were poor, technically. Regular TLD audit seems to be a good way of eliminating errors in basic clinical dosimetry. (author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
85
Journal Issue
1-4
Journal Page Range
p. 393-396
ISSN
0144-8420
CODEN
RPDODE

Conference

Title
12. international conference on solid state dosimetry. Part 2
Dates
5-10 Jul 1998
Place
Burgos (Spain)