Published August 1, 2019 | Version v1
Journal article

Verification of stereotactic cranial radiotherapy treatments with MR-based gel dosimeters: practical aspects

  • 1. Joint Department of Physics, The Institute of Cancer Research and The Royal Marsden NHS Foundation Trust, London SM2 5NG (United Kingdom)
  • 2. CR-UK and EPSRC Cancer Imaging Centre, Royal Marsden NHS Foundation Trust, London SM2 5NG (United Kingdom)
  • 3. Radiotherapy Physics department, Georges Pompidou European Hospital, Paris (France)
  • 4. Radiotherapy Physics, Royal Marsden NHS Foundation Trust, London, SW3 6JJ (United Kingdom)

Description

Image Guided Radiotherapy (IGRT) systems allow delivering high doses of radiation to a tumour with great precision and accuracy. In this work, we use patient-specific head MR-based gel dosimeters as an end-to-end test to commission IGRT delivery systems, for stereotactic cranial radiotherapy, and investigate possible sources of errors associated with practical aspects of the measurements. A CT scan of a patient is used to 3D print the shape of the cranial bones to create head phantoms with MR-based gel inside. Each phantom was used as a patient, following the radiotherapy workflow which includes: taking a CT scan of the phantom, calculating the dose distribution using a treatment planning system (TPS) and delivering the radiation calculated by the TPS (to target three different lesions: Big, Medium and Small). Each phantom was then scanned on an MR scanner to obtain a T2 map (linear with dose), which was then rigidly registered with the CT scan of the phantom based on the phantom structures. Good agreement was obtained between the normalized R2 (=1/T2) values and TPS simulations. The relationship between R2 values and dose was investigated based on 3D regions including each lesion. Good linearity was found, but different R2-dose values relationship was obtained depending on the selected regions. The impact of applying a registration based on each lesion (and not the phantom structures) was also tested. Results show that the registration has an impact on the relationship between R2 values and dose, and although absolute dose measurements are still not possible with these MR-based gel dosimeters, they provide very detailed geometrical 3D information to validate IGRT radiotherapy treatments with high level of accuracy. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/1305/1/012025

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
1305
Journal Issue
1
Journal Page Range
[7 p.]
ISSN
1742-6596

Conference

Title
10. International Conference on 3D Radiation Dosimetry
Acronym
IC3DDose
Dates
16-19 Sep 2018
Place
Kunshan (China)