Evaluation of spatial resolution on fast spin-echo images. Effect of echo train length and interecho spacing
- 1. Kyoto City Hospital (Japan)
Description
Fast spin echo (FSE) imaging technique has been indispensable in clinical MRI. This technique reduces acquisition time compared with conventional spin echo sequences. However, this technique introduces new factors, such as echo train length (ETL) and interecho spacing and is susceptible to a new set of artifacts, such as image blurring secondary to nonuniform sampling of K-space. Therefore, the effects of spatial resolution of FSE images on echo train length and interecho spacing were measured using the final MTF. FSE images of the chart phantom with some ETLs and bandwidth were exposed on laser films and measured by micro-densitometer in the phase-encoding direction. Using the final MTF method, the FSE images were evaluated for spatial resolution. Increasing ETL reduced spatial resolution on FSE images. The proton-density-weighted images were more conspicuous. Decreasing the bandwidth reduced spatial resolution on FSE images because of increasing interecho spacing. (author)
Additional details
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi
- Journal Volume
- 53
- Journal Issue
- 5
- Journal Page Range
- p. 590-594.
- ISSN
- 0369-4305
- CODEN
- NIPHAP
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 28055063
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- IMAGES; NMR IMAGING; PERFORMANCE TESTING; SIGNALS; SPATIAL RESOLUTION; SPIN ECHO; SPIN-SPIN RELAXATION; TRANSFER MATRIX METHOD
- Descriptors DEC
- CALCULATION METHODS; DIAGNOSTIC TECHNIQUES; RELAXATION; RESOLUTION; TESTING