Charge-injection-device performance in the high-energy-neutron environment of laser-fusion experiments
Creators
- 1. Laboratory for Laser Energetics, University of Rochester, 250 East River Road, Rochester, New York 14623 (United States)
Description
Charge-injection devices (CIDs) are being used to image x rays in laser-fusion experiments on the University of Rochester's OMEGA Laser System. The CID cameras are routinely used up to the maximum neutron yields generated (∼1014 DT). The detectors are deployed in x-ray pinhole cameras and Kirkpatrick-Baez microscopes. The neutron fluences ranged from ∼107 to ∼109 neutrons/cm2 and useful x-ray images were obtained even at the highest fluences. It is intended to use CID cameras at the National Ignition Facility (NIF) as a supporting means of recording x-ray images. The results of this work predict that x-ray images should be obtainable on the NIF at yields up to ∼1015, depending on distance and shielding.
Additional details
Identifiers
- DOI
- 10.1063/1.3471960;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 81
- Journal Issue
- 10
- Journal Page Range
- p. 10E503-10E503.4
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44016264
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
- Descriptors DEI
- BEAM INJECTION; CAMERAS; CHARGED PARTICLES; FAST NEUTRONS; IMAGES; LASER FUSION REACTORS; LASERS; MICROSCOPES; NEUTRON FLUENCE; OMEGA FACILITY; US NATIONAL IGNITION FACILITY; X RADIATION; X-RAY DETECTION
- Descriptors DEC
- BARYONS; DETECTION; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; FERMIONS; HADRONS; IONIZING RADIATIONS; NEUTRONS; NUCLEONS; RADIATION DETECTION; RADIATIONS; THERMONUCLEAR REACTORS
Optional Information
- Notes
- (c) 2010 American Institute of Physics