Fundamental study on a long-duration flash X-ray generator with a surface-discharge triode
Creators
- 1. Iwate Medical Univ., Morioka (Japan). School of Medicine
Description
Fundamental studies on a long-duration flash X-ray generator are described. This generator consisted of the following components: a high-voltage power supply with a maximum voltage of 100 kV, an energy-storage condenser of 500 nF, a main discharge condenser of 10 nF, a turbo molecular pump, a thyratron pulser as a trigger device, and a surface-discharge triode. The effective pulse width was less than 30 μs, and the X-ray intensity approximately had a value of 0.6 μC/kg at 1.0 m per pulse with a charged voltage of 60 kV. The maximum tube voltage was equivalent to the initial charged voltage of the condenser, and the peak tube current was less than 40 A. With this generator, we could obtain stable X-ray intensity maximized by preventing damped oscillations of the tube voltage and current. (author)
Additional details
Publishing Information
- Journal Title
- Japanese Journal of Applied Physics. Part 1, Regular Papers and Short Notes
- Journal Volume
- 33
- Journal Issue
- 7 A
- Journal Page Range
- p. 4146-4151.
- ISSN
- 0021-4922
- CODEN
- JAPNDE
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 26040860
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; ELECTRIC POTENTIAL; PULSE TECHNIQUES; TRIODE TUBES; X-RAY SOURCES; X-RAY TUBES
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; ELECTRON TUBES; EQUIPMENT; MEDICINE; RADIATION SOURCES; X-RAY EQUIPMENT