Published September 2007 | Version v1
Journal article

Accurate and precise determination of small quantity uranium by means of automatic potentiometric titration

  • 1. China Inst. of Atomic Energy, Beijing (China)

Description

For high radioactivity level of dissolved solution of spent fuel and the solution of uranium product, radioactive hazard must be considered and reduced as low as possible during accurate determination of uranium. In this work automatic potentiometric titration was applied and the sample only 10 mg of uranium contained was taken in order to reduce the harm of analyzer suffered from the radioactivity. RSD<0.06%, at the same time the result can be corrected for more reliable and accurate measurement. The determination method can effectively reduce the harm of analyzer suffered from the radioactivity, and meets the requirement of reliable accurate measurement of uranium. (authors)

Additional details

Publishing Information

Journal Title
Atomic Energy Science and Technology
Journal Volume
41
Journal Issue
5
Journal Page Range
p. 546-549
ISSN
1000-6931

Optional Information

Notes
1 fig., 4 tabs., 6 refs.