Published June 11, 2016 | Version v1
Journal article

A measurement method of a detector response function for monochromatic electrons based on the Compton scattering

  • 1. St.Petersburg Nuclear Physics Institute, National Research Center "Kurchatov Institute", Gatchina 188300 (Russian Federation)
  • 2. Saint-Petersburg State University, Universitetskaja nab. 7/9, Saint-Petersburg 199034 (Russian Federation)
  • 3. GranSasso Science Institute, INFN, L'Aquila (AQ) I-67100 (Italy)

Description

In this paper we present a method of scintillation detector energy calibration using the gamma-rays. The technique is based on the Compton scattering of gamma-rays in a scintillation detector and subsequent photoelectric absorption of the scattered photon in the Ge-detector. The novelty of this method is that the source of gamma rays, the germanium and scintillation detectors are immediately arranged adjacent to each other. The method presents an effective solution for the detectors consisting of a low atomic number materials, when the ratio between Compton effect and photoelectric absorption is large and the mean path of gamma-rays is comparable to the size of the detector. The technique can be used for the precision measurements of the scintillator light yield dependence on the electron energy.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2016.03.054

Additional details

Identifiers

DOI
10.1016/j.nima.2016.03.054;
arXiv
arXiv:1604.04128v1;
PII
S0168-9002(16)30071-7;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
821
Journal Page Range
p. 13-16
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2016 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.