Published March 2010 | Version v1
Journal article

Determination of uranium in environment samples by ICP-AES with solid phase extraction separation into D293 resin

  • 1. Faculty of Chemistry Biology and Materials Science, East China Institute of Technology, Fuzhou (China)

Description

The optimum conditions for separation and preconcentration of uranium are studied in this paper. Uranium is adsorbed by D293 resins at pH 1.5, and eluted completely using 10 mL of 0.9 mol/L NH4NO3-0.1 mol/L HNO3, then determined by ICP-AES.The detection limit, RSD and related error are of 0.05 μg/mL, 0.8% ∼ 2.9% and 0.4% ∼ 9.8%, respectively. This method is simple, quick and accurate. It can be applied to determine uranium in environment samples with satisfactory results. (authors)

Additional details

Publishing Information

Journal Title
Journal of East China Institute of Technology. Natural Science
Journal Volume
33
Journal Issue
1
Journal Page Range
p. 71-74
ISSN
1674-3504

Optional Information

Notes
4 figs., 2 tabs., 10 refs.