The validation of tomotherapy dose calculations in low-density lung media
Creators
- 1. Department of Radiation Oncology, Washington University School of Medicine, St. Louis, MO 63110 (United States)
Description
The dose-calculation accuracy of the tomotherapy Hi-Art II (registered) (Tomotherapy, Inc., Madison, WI) treatment planning system (TPS) in the presence of low-density lung media was investigated. In this evaluation, a custom-designed heterogeneous phantom mimicking the mediastinum geometry was used. Gammex LN300 and balsa wood were selected as two lung-equivalent materials with different densities. Film analysis and ionization chamber measurements were performed. Treatment plans for esophageal cancers were used in the evaluation. The agreement between the dose calculated by the TPS and the dose measured via ionization chambers was, in most cases, within 0.8%. Gamma analysis using 3% and 3 mm criteria for radiochromic film dosimetry showed that 98% and 95% of the measured dose distribution had passing gamma values ≤1 for LN300 and balsa wood, respectively. For a homogeneous water-equivalent phantom, 95% of the points passed the gamma test. It was found that for the interface between the low-density medium and water-equivalent medium, the TPS calculated the dose distribution within acceptable limits. The phantom developed for this work enabled detailed quality-assurance testing under realistic conditions with heterogeneous media.
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/54/8/004Additional details
Identifiers
- DOI
- 10.1088/0031-9155/54/8/004;
- PII
- S0031-9155(09)75309-6;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 54
- Journal Issue
- 8
- Journal Page Range
- p. 2315-2322
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41012387
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; COMPUTERIZED TOMOGRAPHY; CT-GUIDED RADIOTHERAPY; DENSITY; FILM DOSIMETRY; IONIZATION CHAMBERS; LUNGS; MEDIASTINUM; PHANTOMS; PLANNING; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; VALIDATION; WATER
- Descriptors DEC
- BODY; CHEST; DIAGNOSTIC TECHNIQUES; DOSES; DOSIMETRY; HYDROGEN COMPOUNDS; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; OXYGEN COMPOUNDS; PHYSICAL PROPERTIES; RADIATION DETECTORS; RADIOLOGY; RADIOTHERAPY; RESPIRATORY SYSTEM; STRUCTURAL MODELS; TESTING; THERAPY; TOMOGRAPHY