Published 1976 | Version v1
Book

Low energy x-ray sources utilized at NBS for detector calibration

  • 1. National Bureau of Standards, Washington, DC

Description

Low energy x-ray sources (1 to 10 keV) have been produced utilizing dc and pulsed electron beams to excite the characteristic K-line radiation of elemental targets. The steady state x-ray sources are continuously variable in intensity from 1012 K x rays per sr . s. A parallel plate ionization chamber was calibrated in the steady state sources and used with Ross filter pairs to measure the 10 ns pulsed source K x-ray fluence. Typical measured peak x-ray flux intensities are 10-18 to 1019 K x rays per sr . s. The pulsed source contains four identical ports for simultaneous measurements of the K x-ray fluence by the ionization chamber and detectors being calibrated. The steady state and pulsed K x-ray sources, which differ in flux density by 10 orders of magnitude, have been utilized to determine the responses of a photoelectric diode, a scintillator-vacuum photodiode combination, a silicon PIN vacuum photodiode combination, and a silicon PIN diode

Additional details

Publishing Information

Publisher
Institute of Electrical and Electronics Engineers, Inc.
Imprint Place
New York
Imprint Title
Proceedings of the fourth conference on the scientific and industrial applications of small accelerators
Journal Page Range
p. 195-202.

Conference

Title
4. annual conference on the use of small accelerators.
Dates
25 Oct 1976.
Place
Denton, TX, USA.

Optional Information

Notes
Imprint:See CONF-761059--.