Published May 2010 | Version v1
Journal article

Design of magnetic analysis system for magnetic proton recoil spectrometer

  • 1. Institute of Nuclear Physics and Chemistry, Mianyang (China)
  • 2. Department of Engineering Physics, Tsinghua University, Beijing (China)

Description

Magnetic proton recoil (MPR) spectrometer is a novel diagnostic instrument with high performance for measurements of the neutron spectra from inertial confinement fusion (ICF) experiments and high power fusion devices. The design of the magnetic analysis system, which is a key part of the compact MPR-type spectrometer, has been completed through two-dimensional beam transport simulations and three-dimensional particle transport simulation. The analysis of the system's parameters and performances was performed, as well as system designs based on preferential principles of energy resolution, detection efficiency, and count rate, respectively. The results indicate that the magnetic analysis system can achieve a detection efficiency of 10-5 ∼ 10-4 level at the resolution range of 1.5% to 3.0% and fulfill the design goals of the compact MPR spectrometer. (authors)

Additional details

Publishing Information

Journal Title
High Power Laser and Particle Beams
Journal Volume
22
Journal Issue
5
Journal Page Range
p. 1141-1144
ISSN
1001-4322

Optional Information

Notes
7 figs., 2 tab., 10 refs.