Evaluation of gamma contribution in a neutron calibration field of Am−Be241 source

  • 1. Institute for Nuclear Science and Technology – VINATOM, 179 Hoang Quoc Viet, Hanoi (Viet Nam)
  • 2. Institute of Fundamental and Applied Sciences, Duy Tan University, 3 Quang Trung, Da Nang (Viet Nam)
  • 3. Vietnam Agency for Radiation and Nuclear Safety, 113 Tran Duy Hung, Hanoi (Viet Nam)

Description

The paper presents the evaluation of gamma contribution in a neutron calibration field of a 241AmBe source. The characterization of gamma flux spectra and gamma ambient dose equivalent rates has been performed using a portable NaI(Tl) gamma spectrometer. For obtaining the gamma ambient dose equivalent rates from the measured spectra, two approaches were applied based on the applications of the G(E) function and the ICRP 74 conversion factors, respectively. Comparison of the gamma ambient dose equivalent rates obtained from the two approaches shows an agreement within 5%. The gamma contribution in the neutron calibration field is evaluated about 2.3–3.3% of the total neutron ambient dose equivalent rate in the distance range of 100–250 cm.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2018.01.001

Additional details

Identifiers

DOI
10.1016/j.apradiso.2018.01.001;
PII
S0969804317310709;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
133
Journal Page Range
p. 117-120
ISSN
0969-8043
CODEN
ARISEF

Optional Information

Copyright
Copyright (c) 2017 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.